  • Environmental Express MicroBlock-R System with Ammonia Glassware; 240V

    C5201-240 MicroBlock™-V, Block Only, 240 VAC

    (adjustable digital temperature control with LCD readout and timer, Includes a flow meter to monitor cooling water rates)

    Materials of Construction PTFE-coated aluminum block, powder-coated steel housing
    Power 240 VAC, 7.5 amps
    Power (Hz) 50

    Item#   Description
    C5231-240            MicroBlock R, 240v, Block Only
    C5214B   Ammonia Reaction Tube, each
    C5219   Ammonia Distillation Head, each
    C5210   Cold Finger for Distillation Glassware, each
    C5228   Receiver Tube for Ammonia/Phenol Distillation, each
    C5226   Outlet Tube for Ammonia Distillation, each
    C5213   Ammonia Pierced Cap, 10pk
    C5218   PTFE-Faced Septas for Distillation Cap, 20pk
    C5225   Silicon Sealing Ring, 10pk

    7.0 PROCEDURE
    7.1 Preparation of distillation glassware. Steam out the distillation glassware using 50mL reagent water with a pinch of
    boiling chips until no trace of ammonia can be detected.
    7.2 If residual chlorine is present add sodium thiosulfate or sodium sulfite equivalent to the chlorine level.
    7.3 To a 50mL sample add 1N NaOH drop-wise with mixing and adjust to pH 9.5.
    7.4 Transfer sample to the Reaction Tube, skirted, and add 2.5mL of borate buffer followed by a pinch of boiling chips.
    7.5 To the Receiver Tube add 5-10mL of 20g/L Boric Acid or 0.2N H2SO4. If using the Snip & Pour concentrate pour the
    entire contents of the tube in the Reaction Tube. The tip of the long stem must be below the level of boric acid or
    sulfuric acid.
    7.6 Repeat for all samples to be distilled up to a maximum of 10.
    7.7 Assemble the distillation glassware setup as shown in Figure 3.
    7.8 Turn on the tap water to cool the Cold Finger after ensuring all tubing connections are tight. Check for any leaks.
    7.9 Turn the power ON and set heater block temperature to 190°C and the timer to 60 minutes.
    7.10 Collect a minimum of 30mL of distillate.
    7.11 Turn power OFF.
    7.12 Unscrew the cap holding the Long Stem in the receiver tube and remove from all distillation positions to prevent
    any suck back of the distilled sample. When the Long Stem is disconnected, gently let it rest at the bottom of the
    Receiver Tube.